MV-401 Mass Flow Controller
                            
                    Mass Flow Controller for gases with pressure compensation and integrated display
- Direct thermal mass flow measurement
- High accuracy
- 4 pre-installed gases
- Bright, graphical OLED display
- Digital or analog output; alarm and counter functions
- High precision needle valve with pressure compensation
Valve
        0 .. 50°C
        Pre-installed ranges
| Ranges | Flow range (relative Air) | 
|---|---|
| Range 1 | 20 mln/min .. 200 mln/min (sccm) | 
| Range 2 | 20 mln/min .. 100 mln/min (sccm) | 
| Range 3 | 20 mln/min .. 50 mln/min (sccm) | 
Pre-installed gases
| Gases | min. range .. max. range | 
|---|---|
| Air, N2, O2 and CO | 20 .. 200 mln/min (sccm) | 
Features of MV-401 Mass Flow Controller
- Clear indication in:
 - actual flow rate (bar graph and value)
 - desired flow units
 - type of gas
 - totalised flow
- Bright, wide-angle OLED display
- Free of parallax error
- Virtually independent of temperature and
 pressure variations
- Low pressure drop
- Wide flow ranges
- Fast response
- High accuracy
- Electronic output:
 - Analog (0 .. 5 Vdc)
 - RS-232 interface (with Bronkhorst FlowSuite 2)
 - Modbus-RTU/ASCII interface
- Configurable password protection
- Alarm and counter functions
- Relays
- Multi Gas / Multi Range
- 4 Pre-installed gases
- Membrane operated valve
- High resolution needle valve
- Digitally calibrated
- Free software tools
Scope of delivery
- MASS-VIEW flow controller
- packing box
- 9.17.054 Short form manual
- (optional) fittings
- (optional) power supply
Technical specifications
Measurement / control system
Flow range 
                
                    
                    
                    
                    
                    20 mln/min .. 200 mln/min (sccm)
Mass flow units ln/min and SLM refer to normal operating conditions, i.e. 0°C (32°F) and 1013 mbar (14.7 psi)
            Mass flow units ln/min and SLM refer to normal operating conditions, i.e. 0°C (32°F) and 1013 mbar (14.7 psi)
Accuracy
                
                    
                    
                    
                    
                    ± 2% RD for flow > 50% of max. capacity
± (1% RD + 0.5% FS) on lower flows
Technical specifications are based on Air at maximum FS
            ± (1% RD + 0.5% FS) on lower flows
Technical specifications are based on Air at maximum FS
Repeatability
                
                    
                    
                    
                    
                    < 0.2% FS typical
                
            Turndown ratio
                
                    
                    
                    
                    
                    up to 1:10
                
            Temperature rating
                
                    
                    
                    
                    
                    0 … 50 (32 … 122°F)
                
            Operating pressure
                
                    
                    
                    
                    
                    0 … 10 bar(g) / 0 … 150 psi(g)
                
            Pressure coefficient
                
                    
                    
                    
                    
                    0.2% / bar typical at Air
                
            Pre-installed gases
                
                    
                    
                    
                    
                    Air, N2, O2, CO
                
            Standard calibration gas
                
                    
                    
                    
                    
                    Air, other gases are converted using our Fluidat® conversion model which will introduce extra inaccuracy
                
            Rangeability
                
                    
                    
                    
                    
                    up to 1:100
                
            Response time (sensor)
                
                    
                    
                    
                    
                    (t63%) 	2 seconds
                
            Temperature coefficient
                
                    
                    
                    
                    
                    Zero: <0.1%FS/°C, Span: <0.2%RD/°C
                
            Leak integrity, outboard
                
                    
                    
                    
                    
                    tested < 2 x 10-9 mbar l/s
                
            Attitude sensitivity
                
                    
                    
                    
                    
                    < 0,1 % FS
                
            Mechanical parts
Material (wetted parts)
                
                    
                    
                    
                    
                    Meter: aluminium, Viton (FPM)
Needle Valve: Fluorosint, Brass, Buna N
Controller: membrane Fiber-reinforced nitrile
            Needle Valve: Fluorosint, Brass, Buna N
Controller: membrane Fiber-reinforced nitrile
Test pressure
                
                    
                    
                    
                    
                    21 bar(a) / 300 psi(a)
                
            Min. ΔP
                
                    
                    
                    
                    
                    1 bar(d)
                
            Max. ΔP
                
                    
                    
                    
                    
                    7 bar (d)
                
            Process connections
                
                    
                    
                    
                    
                    G 1/4" BSPP female
(compression fittings optional)
            (compression fittings optional)
Sealing material
                
                    
                    
                    
                    
                    Viton®, PTFE, Fluorosint®, Buna N;
Controller membrane: fiber-reinforced nitrile
            Controller membrane: fiber-reinforced nitrile
Weight
                
                    
                    
                    
                    
                    1,4 kg
                
            Ingress protection
                
                    
                    
                    
                    
                    IP40
                
            Control Accuracy
                
                    
                    
                    
                    
                    typical <1%/bar @ N2
                
            Electrical properties
Power supply
                
                    
                    
                    
                    
                    +15 … 24 Vdc (+/- 10%)
                
            Max. power consumption
                
                    
                    
                    
                    
                    approx 135 mA
                
            Analog output
                
                    
                    
                    
                    
                    0 … 5 Vdc
                
            EMC
                
                    
                    
                    
                    
                    EU declaration
                
            Digital Communication
                
                    
                    
                    
                    
                    RS232 / RS485 (Modbus RTU/ASCII)
                
            Min. And max. Relay contacts
                
                    
                    
                    
                    
                    switching current 0.5 A, 24 Vdc
one side grounded (0 Vdc power)
            one side grounded (0 Vdc power)
Electrical connection
                
                    
                    
                    
                    
                    8-pin RJ-45 modular jack
                
            Downloads
Brochure & Certificate
Dimensional drawing
Stepfile
Applications
The fields of application for MASS-VIEW® mass flow controllers are diverse:
- Burner Control (furnace construction)
- Welding (welding gas monitoring)
- Leak measurements (quality, environment)
- Coating (equipment construction)
- Regulation of gaseous atmospheres (biotechnology)
- Measurements of gas consumption (hospitals)
- Test equipment (production maintenance)
- Local preparation of a gas mixture
- Flow rate monitoring (laboratories)
- Cutting (steel sheets)
- Cost centre billing
- Analytical equipment
- Aeration / sparging (food products, ice cream / edible oils
- Blanketing (food)
- Fermentation (food, biotechnology, pharma)
Frequently Asked Questions
- Is there a pressure drop chart of the MV-400 series controller available?
- Accuracy & Repeatability Flow meters
- Menu options on display
- How to connect several MASS-VIEWs (5, 6, 7) to a RS-485 bus system?
- Is MASS-VIEW compatible with Oxygen (degreased)?
- How to connect MASS-VIEW® using RS-232 to a computer?
- What is Aeration?
- What do ln/min, Is/min, slm and sccm stand for?
- What is the purpose and operation of FlowDDE?
- Which flow units are available to select?
More questions on our FAQ page
 
                                                 
                                                

 
                        