How to set mass counter units in MASS-VIEW?

It is currently not possible to set the counter units in the MASS-VIEW to use “mass” units (ug, mg, g, kg) using the menu. There is a “work around” by setting this parameter using a Windows PC with RS-232 or USB.

This is the procedure how to do this.

Your Question:  Customer can not read the Counter unit in mass flow

Solution:  Change the sensortype of the instrument using FlowDDE ( parameter 22 from 131 into 130).

Action to take:

  • Change parameter 22 from value 131 to value 130:
  • Connect instrument to a Serial Comm port of a PC.
  • Install FlowDDE Software
  • Make FlowDDE connection with the instrument and write/read values by FlowDDE server menu “Test FLOW-BUS and DDE F6 “.
  • Select the parameter number by the dropdown menu or by entering the number in the box, select read or write as described.
  • change parameter 22 from value 131 to value 130:–
  • Parameter 7 (initreset)     –> Write value 64 (Soft will enable writing parameter 22).
  • Parameter 7 (initreset)     –> Read parameter and check value.
  • Parameter 22 (Sensortype)–> Write value 130.
  • Parameter 22 (Sensortype)–> Read parameter and check value.
  • Parameter 7 (initreset)     –> Write value 82 (NormalRunningMode).
  • Parameter 7 (initreset)     –> Read parameter and check value.

Now the Counter value is available in the mass units